EN | EU | ES

EMC semi-anechoic chamber. EMC laboratory

Keywords: 
EMC, EMI, radiated emissions, radiated immunity.
Research Area: 
Physical Sciences
Engineering & Technology
Location: 

IKERLAN, S.Coop.
Pº. J. Mª. Arizmendiarrieta, 2
20500 Arrasate-Mondragón
Tél.: +34 943712400
Fax:.+34 943796944

Contact: 

Roberto Moreno. Embedded Systems Area.Hardware Platform team.
rmoreno@ikerlan.es

Available for external users?: 
Yes
Access protocol: 

The access protoco shall be defined according to the case

Description: 

Semi anechoic chamber with hybrid absorber and ferrite tile, 7,23 m x 4,83 m x 3,055 (h) m (LxWxH) Internal shield, 3 meter measurement.
Radiated emission 30 MHz-1 GHz Pre-Compliant QZ 1,2 x 2,0m. Test volume is cylinder, height is referenced to floor. 30-200 MHz NSA ± 6 dB. 200-1000 MHz NSA ± 4dB.
Radiated immunity 80 MHz-1 GHz Compliant TP 1,5 x 1,5 m.

Equipment:
• Electromagnetic field emission measurement equipment: EMI receiver, antennas, spectrum analyser, LISN, field probes, measurement software.
• Electromagnetic field immunity equipment: amplifiers, field probes, antennas, measurement software, transient generators, couplers, EM clamps.
• Passive components, filters, ferrites.

Services: 

• Pre-certification test.
• Radiated emission test, 3m distance. EN55022, EN55011, etc.
• Radiated immunity test, to 20v/m. EN61000-4-3, 80Mhz – 4,2Ghz.
• Conducted emission test. LISNs. EN55011, EN55022, EN55014-1, etc.
• EN61000-4-2, EN61000-4-4, EN61000-4-5, EN61000-4-6, EN61000-4-8, EN61000-4-9, EN61000-4-10, EN61000-4-11, EN61000-4-13.
• Flicker and harmonics test.
• EMC advice and guidance services.

Other relevant information: 
Wide experience (more than 20 years) in test, requirement analysis and development of EMC solutions at concept level, design level and system level. A team of 4 engineers / researchers working full-time on the EMC and electrical safety development and qualification of products. Close to designers EMC consulting. EMC team, design team working in laboratory